What are you looking for?
The allrounder: Particle-rich, salt-containing, or very pure samples? This combustion analyzer is optimized for high throughput and offers the perfect balance between samples of all kinds.
TOC/TNb analysis for samples with complex matrices and particle contents
The versatile devices of the multi N/C 3300 series are suited for a wide spectrum of applications such as particle-rich surface water or wastewater, as well as drinking water or ultrapure water. Equipped with special loop injection technology, these devices offer especially high reproducibility for samples containing particles. Thanks to the high injection volumes (1000 – 3000 µL), they also offer exceptional sensitivity.
Detector modules for simultaneous TNb determination and accessories for analysis of solids can be pre-configured or retrofitted. The N/C 3300 HS provides even higher reproducibility and sensitivity in the trace range, for example, when analyzing pharmaceutical ultrapure water. An optional solids module can be added for direct swab combustion in the context of TOC cleaning validation.
- Complete dosing of particulate samples without carryover thanks to loop injection technology with reverse rinse function
- Measurement of undiluted samples using the wide-range focus radiation NDIR detector
- Automatic acidification and purging as well as sample homogenization
- High NPOC sample throughput by parallel purging and measurement
- New intuitive software with live result display and 21 CFR Part 11 compliant pharma module
- Significant time savings as a result of long-term stable calibrations
- External needle rinsing for even better repeatability and carry-over control
- Optional kit to measure high-salt samples with salt trap in the combustion tube and extended maintenance intervals
- Five-year warranty for the focus radiation NDIR detector and furnace technology
Brochures
| Title | Language | Info |
|---|---|---|
| Flyer Calibration Standard Offering for TOC, AAS, and ICP Devices (EN) | en | PDF, 492 KB |
| Flyer Kalibrierstandards für TOC-, AAS- und ICP-Geräte (DE) | de | PDF, 633 KB |
| Brochure multi N/C x300 Series (DE) | de | PDF, 1 MB |
| Brochure multi N/C x300 Series (EN) | en | PDF, 1 MB |
| Brochure multi N/C x300 Series (KOR) | ko | PDF, 2 MB |
| Flyer multi N/C Software (DE) | de | PDF, 259 KB |
| Flyer multi N/C Software (EN) | en | PDF, 249 KB |
| Flyer Chemical Recycling of Plastics: Analytical Solutions for Crucial Quality Gates (EN) | en | PDF, 301 KB |
| Brochure multi N/C x300 Series (FR) | fr | PDF, 6 MB |
| Flyer Salt Kit for multi N/C 3300 (EN) | en | PDF, 1 MB |
| Flyer TOC vs. COD/BOD5 – Switch and Save (EN) | en | PDF, 1 MB |
| Flyer TOCnology made for you (EN) | en | PDF, 495 KB |
| Flyer TOCnology made for Pharma EN | en | PDF, 893 KB |
TOC/TNb analysis for samples with complex matrices and particle contents
The versatile devices of the multi N/C 3300 series are suited for a wide spectrum of applications such as particle-rich surface water or wastewater, as well as drinking water or ultrapure water. Equipped with special loop injection technology, these devices offer especially high reproducibility for samples containing particles. Thanks to the high injection volumes (1000 – 3000 µL), they also offer exceptional sensitivity.
Detector modules for simultaneous TNb determination and accessories for analysis of solids can be pre-configured or retrofitted. The N/C 3300 HS provides even higher reproducibility and sensitivity in the trace range, for example, when analyzing pharmaceutical ultrapure water. An optional solids module can be added for direct swab combustion in the context of TOC cleaning validation.
- Complete dosing of particulate samples without carryover thanks to loop injection technology with reverse rinse function
- Measurement of undiluted samples using the wide-range focus radiation NDIR detector
- Automatic acidification and purging as well as sample homogenization
- High NPOC sample throughput by parallel purging and measurement
- New intuitive software with live result display and 21 CFR Part 11 compliant pharma module
- Significant time savings as a result of long-term stable calibrations
- External needle rinsing for even better repeatability and carry-over control
- Optional kit to measure high-salt samples with salt trap in the combustion tube and extended maintenance intervals
- Five-year warranty for the focus radiation NDIR detector and furnace technology
Brochures
| Title | Language | Info |
|---|---|---|
| Flyer Calibration Standard Offering for TOC, AAS, and ICP Devices (EN) | en | PDF, 492 KB |
| Flyer Kalibrierstandards für TOC-, AAS- und ICP-Geräte (DE) | de | PDF, 633 KB |
| Brochure multi N/C x300 Series (DE) | de | PDF, 1 MB |
| Brochure multi N/C x300 Series (EN) | en | PDF, 1 MB |
| Brochure multi N/C x300 Series (KOR) | ko | PDF, 2 MB |
| Flyer multi N/C Software (DE) | de | PDF, 259 KB |
| Flyer multi N/C Software (EN) | en | PDF, 249 KB |
| Flyer Chemical Recycling of Plastics: Analytical Solutions for Crucial Quality Gates (EN) | en | PDF, 301 KB |
| Brochure multi N/C x300 Series (FR) | fr | PDF, 6 MB |
| Flyer Salt Kit for multi N/C 3300 (EN) | en | PDF, 1 MB |
| Flyer TOC vs. COD/BOD5 – Switch and Save (EN) | en | PDF, 1 MB |
| Flyer TOCnology made for you (EN) | en | PDF, 495 KB |
| Flyer TOCnology made for Pharma EN | en | PDF, 893 KB |
TOC/TNb analysis for samples with complex matrices and particle contents
The versatile devices of the multi N/C 3300 series are suited for a wide spectrum of applications such as particle-rich surface water or wastewater, as well as drinking water or ultrapure water. Equipped with special loop injection technology, these devices offer especially high reproducibility for samples containing particles. Thanks to the high injection volumes (1000 – 3000 µL), they also offer exceptional sensitivity.
Detector modules for simultaneous TNb determination and accessories for analysis of solids can be pre-configured or retrofitted. The N/C 3300 HS provides even higher reproducibility and sensitivity in the trace range, for example, when analyzing pharmaceutical ultrapure water. An optional solids module can be added for direct swab combustion in the context of TOC cleaning validation.
- Complete dosing of particulate samples without carryover thanks to loop injection technology with reverse rinse function
- Measurement of undiluted samples using the wide-range focus radiation NDIR detector
- Automatic acidification and purging as well as sample homogenization
- High NPOC sample throughput by parallel purging and measurement
- New intuitive software with live result display and 21 CFR Part 11 compliant pharma module
- Significant time savings as a result of long-term stable calibrations
- External needle rinsing for even better repeatability and carry-over control
- Optional kit to measure high-salt samples with salt trap in the combustion tube and extended maintenance intervals
- Five-year warranty for the focus radiation NDIR detector and furnace technology
Brochures
| Title | Language | Info |
|---|---|---|
| Flyer Calibration Standard Offering for TOC, AAS, and ICP Devices (EN) | en | PDF, 492 KB |
| Flyer Kalibrierstandards für TOC-, AAS- und ICP-Geräte (DE) | de | PDF, 633 KB |
| Brochure multi N/C x300 Series (DE) | de | PDF, 1 MB |
| Brochure multi N/C x300 Series (EN) | en | PDF, 1 MB |
| Brochure multi N/C x300 Series (KOR) | ko | PDF, 2 MB |
| Flyer multi N/C Software (DE) | de | PDF, 259 KB |
| Flyer multi N/C Software (EN) | en | PDF, 249 KB |
| Flyer Chemical Recycling of Plastics: Analytical Solutions for Crucial Quality Gates (EN) | en | PDF, 301 KB |
| Brochure multi N/C x300 Series (FR) | fr | PDF, 6 MB |
| Flyer Salt Kit for multi N/C 3300 (EN) | en | PDF, 1 MB |
| Flyer TOC vs. COD/BOD5 – Switch and Save (EN) | en | PDF, 1 MB |
| Flyer TOCnology made for you (EN) | en | PDF, 495 KB |
| Flyer TOCnology made for Pharma EN | en | PDF, 893 KB |
The versatile devices of the multi N/C 3300 series are suited for a wide spectrum of applications such as particle-rich surface water or wastewater, as well as drinking water or ultrapure water. Equipped with special loop injection technology, these devices offer especially high reproducibility for samples containing particles. Thanks to the high injection volumes (1000 – 3000 µL), they also offer exceptional sensitivity.
Detector modules for simultaneous TNb determination and accessories for analysis of solids can be pre-configured or retrofitted. The N/C 3300 HS provides even higher reproducibility and sensitivity in the trace range, for example, when analyzing pharmaceutical ultrapure water. An optional solids module can be added for direct swab combustion in the context of TOC cleaning validation.
Optimally equipped for a multitude of sample matrices
A wide range of applications, varying particle concentrations and high throughput require powerful technologies. Loop injection with intelligent rinsing technology achieves a detection limit of 4 µg/L C, while avoiding sample carryover when handling particle-rich samples. The flow injection from the syringe pump in the multi N/C 3300 HS makes the device ideal for the pharmaceutical analysis of ultrapure water.
The wide-range focus radiation NDIR detector with a measurement range of 0 to 30,000 mg/L TOC enables the measurement of undiluted samples as well as the precise detection of low TOC contents.
The Easy Cal function enables automatic calibration with long-term stability and standard recovery. The underlying VITA flow management system delivers these benefits by effectively compensating for carrier gas fluctuations. Not only does this mean that calibration is only required once a year, it also greatly simplifies the commonly requested equidistant 10-point calibration.
Our automation solutions make sample preparation and analysis more efficient and can be tailored to the needs of our customers. An extensive portfolio of autosamplers enables automatic acidification with parallel purging, magnetic stirring to homogenize samples, external needle rinsing to prevent carryover when handling particle-rich samples, and the processing of polymer septa for TOC analysis in ultrapure water.
TOC solids measurement can be included in the device configuration of a duo system, including automation, or without automation as an additional module. The HT 1300 is available as a solids module for the non-catalytic combustion of solid samples at up to 1800 °C (with additives) in a ceramic combustion tube. Horizontal sample introduction via ceramic boats enables the direct analysis of solid samples weighing up to 3 g without needing to wrap them into foil. Likewise, manual TIC analysis using direct acid digestion is also possible by adding a TIC module.
Upon request, a reliable electrochemical detector (ChD) can be integrated in the device for nitrogen determination. Alternatively, a chemiluminescence detector (CLD) can be added as a separate detector module – either during initial analyzer configuration or later.
Product Variants at a Glance
multi N/C 3300
Its high level of sensitivity, combined with a broad working range, makes this TOC/TNb analyzer a multi-talented device for high throughput laboratories. The device is suited to analyze both high-matrix samples and ultrapure water. Robust loop injection technology with reverse rinse and parallel purging enables rapid operation in NPOC mode.
Solids analysis (optional, manual)
- Non-catalytic high-temperature combustion with HT 1300
- Acid digestion module for TIC analysis in solids
Nitrogen analysis (optional)
- Simultaneous measurement of TOC and TNb
- ChD detector (integrated in the device)
- CLD-300 detector module (can be retrofitted)
Applications
- Surface water monitoring
- Wastewater and eluate analysis
- Analysis of groundwater, raw water and drinking water
- TOC cleaning validation in pharma industry
multi N/C 3300 duo
The highly sensitive TOC/TNb analyzer with its integrated automated solids analysis function is ideal for contract laboratories that handle a wide range of sample types. High sensitivity across a broad working range makes it suitable for both high-matrix and pure water samples, as well as for less homogeneous solid samples. Robust loop injection technology with reverse rinse and parallel purging enables rapid operation in NPOC mode.
Solids analysis
- Automated non-catalytic high-temperature combustion at up to 1800 °C (with additives) in a ceramic combustion tube
- Optional: Acid digestion module for TIC analysis in solids
Nitrogen analysis (optional)
- Simultaneous measurement of TOC and TNb
- ChD detector (integrated in the device)
Applications:
- Surface water monitoring
- Wastewater and eluate analysis
- Analysis of groundwater, raw water and drinking water
- Analysis of soils, contaminated sites, sediment, waste, construction debris
multi N/C 3300 HS
The TOC analyzer for applications at the ultra-trace level. By dosing directly from the syringe pump, the device can utilize high sample volumes of up to 3 mL, thereby achieving the highest sensitivity and repeatability when analyzing the lowest concentrations of TOC and TN.
Solids analysis (optional)
- Catalytic high-temperature combustion of quartz fiber swabs with the swab test module
Nitrogen analysis (optional, can be retrofitted)
- CLD-300 detector module
Applications:
- TOC and TN trace analysis
- Ultrapure water analysis (water for injection – WFI) in pharma industry
- Cleaning validation (last rinse, swab extraction and direct swab combustion)
- Analysis of boiler feed water and condensates in power plants
Optimally equipped for a multitude of sample matrices
A wide range of applications, varying particle concentrations and high throughput require powerful technologies. Loop injection with intelligent rinsing technology achieves a detection limit of 4 µg/L C, while avoiding sample carryover when handling particle-rich samples. The flow injection from the syringe pump in the multi N/C 3300 HS makes the device ideal for the pharmaceutical analysis of ultrapure water.
The wide-range focus radiation NDIR detector with a measurement range of 0 to 30,000 mg/L TOC enables the measurement of undiluted samples as well as the precise detection of low TOC contents.
The Easy Cal function enables automatic calibration with long-term stability and standard recovery. The underlying VITA flow management system delivers these benefits by effectively compensating for carrier gas fluctuations. Not only does this mean that calibration is only required once a year, it also greatly simplifies the commonly requested equidistant 10-point calibration.
Our automation solutions make sample preparation and analysis more efficient and can be tailored to the needs of our customers. An extensive portfolio of autosamplers enables automatic acidification with parallel purging, magnetic stirring to homogenize samples, external needle rinsing to prevent carryover when handling particle-rich samples, and the processing of polymer septa for TOC analysis in ultrapure water.
TOC solids measurement can be included in the device configuration of a duo system, including automation, or without automation as an additional module. The HT 1300 is available as a solids module for the non-catalytic combustion of solid samples at up to 1800 °C (with additives) in a ceramic combustion tube. Horizontal sample introduction via ceramic boats enables the direct analysis of solid samples weighing up to 3 g without needing to wrap them into foil. Likewise, manual TIC analysis using direct acid digestion is also possible by adding a TIC module.
Upon request, a reliable electrochemical detector (ChD) can be integrated in the device for nitrogen determination. Alternatively, a chemiluminescence detector (CLD) can be added as a separate detector module – either during initial analyzer configuration or later.
Product Variants at a Glance
multi N/C 3300
Its high level of sensitivity, combined with a broad working range, makes this TOC/TNb analyzer a multi-talented device for high throughput laboratories. The device is suited to analyze both high-matrix samples and ultrapure water. Robust loop injection technology with reverse rinse and parallel purging enables rapid operation in NPOC mode.
Solids analysis (optional, manual)
- Non-catalytic high-temperature combustion with HT 1300
- Acid digestion module for TIC analysis in solids
Nitrogen analysis (optional)
- Simultaneous measurement of TOC and TNb
- ChD detector (integrated in the device)
- CLD-300 detector module (can be retrofitted)
Applications
- Surface water monitoring
- Wastewater and eluate analysis
- Analysis of groundwater, raw water and drinking water
- TOC cleaning validation in pharma industry
multi N/C 3300 duo
The highly sensitive TOC/TNb analyzer with its integrated automated solids analysis function is ideal for contract laboratories that handle a wide range of sample types. High sensitivity across a broad working range makes it suitable for both high-matrix and pure water samples, as well as for less homogeneous solid samples. Robust loop injection technology with reverse rinse and parallel purging enables rapid operation in NPOC mode.
Solids analysis
- Automated non-catalytic high-temperature combustion at up to 1800 °C (with additives) in a ceramic combustion tube
- Optional: Acid digestion module for TIC analysis in solids
Nitrogen analysis (optional)
- Simultaneous measurement of TOC and TNb
- ChD detector (integrated in the device)
Applications:
- Surface water monitoring
- Wastewater and eluate analysis
- Analysis of groundwater, raw water and drinking water
- Analysis of soils, contaminated sites, sediment, waste, construction debris
multi N/C 3300 HS
The TOC analyzer for applications at the ultra-trace level. By dosing directly from the syringe pump, the device can utilize high sample volumes of up to 3 mL, thereby achieving the highest sensitivity and repeatability when analyzing the lowest concentrations of TOC and TN.
Solids analysis (optional)
- Catalytic high-temperature combustion of quartz fiber swabs with the swab test module
Nitrogen analysis (optional, can be retrofitted)
- CLD-300 detector module
Applications:
- TOC and TN trace analysis
- Ultrapure water analysis (water for injection – WFI) in pharma industry
- Cleaning validation (last rinse, swab extraction and direct swab combustion)
- Analysis of boiler feed water and condensates in power plants
Optimally equipped for a multitude of sample matrices
A wide range of applications, varying particle concentrations and high throughput require powerful technologies. Loop injection with intelligent rinsing technology achieves a detection limit of 4 µg/L C, while avoiding sample carryover when handling particle-rich samples. The flow injection from the syringe pump in the multi N/C 3300 HS makes the device ideal for the pharmaceutical analysis of ultrapure water.
The wide-range focus radiation NDIR detector with a measurement range of 0 to 30,000 mg/L TOC enables the measurement of undiluted samples as well as the precise detection of low TOC contents.
The Easy Cal function enables automatic calibration with long-term stability and standard recovery. The underlying VITA flow management system delivers these benefits by effectively compensating for carrier gas fluctuations. Not only does this mean that calibration is only required once a year, it also greatly simplifies the commonly requested equidistant 10-point calibration.
Our automation solutions make sample preparation and analysis more efficient and can be tailored to the needs of our customers. An extensive portfolio of autosamplers enables automatic acidification with parallel purging, magnetic stirring to homogenize samples, external needle rinsing to prevent carryover when handling particle-rich samples, and the processing of polymer septa for TOC analysis in ultrapure water.
TOC solids measurement can be included in the device configuration of a duo system, including automation, or without automation as an additional module. The HT 1300 is available as a solids module for the non-catalytic combustion of solid samples at up to 1800 °C (with additives) in a ceramic combustion tube. Horizontal sample introduction via ceramic boats enables the direct analysis of solid samples weighing up to 3 g without needing to wrap them into foil. Likewise, manual TIC analysis using direct acid digestion is also possible by adding a TIC module.
Upon request, a reliable electrochemical detector (ChD) can be integrated in the device for nitrogen determination. Alternatively, a chemiluminescence detector (CLD) can be added as a separate detector module – either during initial analyzer configuration or later.
Product Variants at a Glance
multi N/C 3300
Its high level of sensitivity, combined with a broad working range, makes this TOC/TNb analyzer a multi-talented device for high throughput laboratories. The device is suited to analyze both high-matrix samples and ultrapure water. Robust loop injection technology with reverse rinse and parallel purging enables rapid operation in NPOC mode.
Solids analysis (optional, manual)
- Non-catalytic high-temperature combustion with HT 1300
- Acid digestion module for TIC analysis in solids
Nitrogen analysis (optional)
- Simultaneous measurement of TOC and TNb
- ChD detector (integrated in the device)
- CLD-300 detector module (can be retrofitted)
Applications
- Surface water monitoring
- Wastewater and eluate analysis
- Analysis of groundwater, raw water and drinking water
- TOC cleaning validation in pharma industry
multi N/C 3300 duo
The highly sensitive TOC/TNb analyzer with its integrated automated solids analysis function is ideal for contract laboratories that handle a wide range of sample types. High sensitivity across a broad working range makes it suitable for both high-matrix and pure water samples, as well as for less homogeneous solid samples. Robust loop injection technology with reverse rinse and parallel purging enables rapid operation in NPOC mode.
Solids analysis
- Automated non-catalytic high-temperature combustion at up to 1800 °C (with additives) in a ceramic combustion tube
- Optional: Acid digestion module for TIC analysis in solids
Nitrogen analysis (optional)
- Simultaneous measurement of TOC and TNb
- ChD detector (integrated in the device)
Applications:
- Surface water monitoring
- Wastewater and eluate analysis
- Analysis of groundwater, raw water and drinking water
- Analysis of soils, contaminated sites, sediment, waste, construction debris
multi N/C 3300 HS
The TOC analyzer for applications at the ultra-trace level. By dosing directly from the syringe pump, the device can utilize high sample volumes of up to 3 mL, thereby achieving the highest sensitivity and repeatability when analyzing the lowest concentrations of TOC and TN.
Solids analysis (optional)
- Catalytic high-temperature combustion of quartz fiber swabs with the swab test module
Nitrogen analysis (optional, can be retrofitted)
- CLD-300 detector module
Applications:
- TOC and TN trace analysis
- Ultrapure water analysis (water for injection – WFI) in pharma industry
- Cleaning validation (last rinse, swab extraction and direct swab combustion)
- Analysis of boiler feed water and condensates in power plants
Optimally equipped for a multitude of sample matrices
A wide range of applications, varying particle concentrations and high throughput require powerful technologies. Loop injection with intelligent rinsing technology achieves a detection limit of 4 µg/L C, while avoiding sample carryover when handling particle-rich samples. The flow injection from the syringe pump in the multi N/C 3300 HS makes the device ideal for the pharmaceutical analysis of ultrapure water.
The wide-range focus radiation NDIR detector with a measurement range of 0 to 30,000 mg/L TOC enables the measurement of undiluted samples as well as the precise detection of low TOC contents.
The Easy Cal function enables automatic calibration with long-term stability and standard recovery. The underlying VITA flow management system delivers these benefits by effectively compensating for carrier gas fluctuations. Not only does this mean that calibration is only required once a year, it also greatly simplifies the commonly requested equidistant 10-point calibration.
Our automation solutions make sample preparation and analysis more efficient and can be tailored to the needs of our customers. An extensive portfolio of autosamplers enables automatic acidification with parallel purging, magnetic stirring to homogenize samples, external needle rinsing to prevent carryover when handling particle-rich samples, and the processing of polymer septa for TOC analysis in ultrapure water.
TOC solids measurement can be included in the device configuration of a duo system, including automation, or without automation as an additional module. The HT 1300 is available as a solids module for the non-catalytic combustion of solid samples at up to 1800 °C (with additives) in a ceramic combustion tube. Horizontal sample introduction via ceramic boats enables the direct analysis of solid samples weighing up to 3 g without needing to wrap them into foil. Likewise, manual TIC analysis using direct acid digestion is also possible by adding a TIC module.
Upon request, a reliable electrochemical detector (ChD) can be integrated in the device for nitrogen determination. Alternatively, a chemiluminescence detector (CLD) can be added as a separate detector module – either during initial analyzer configuration or later.
Product Variants at a Glance
multi N/C 3300
Its high level of sensitivity, combined with a broad working range, makes this TOC/TNb analyzer a multi-talented device for high throughput laboratories. The device is suited to analyze both high-matrix samples and ultrapure water. Robust loop injection technology with reverse rinse and parallel purging enables rapid operation in NPOC mode.
Solids analysis (optional, manual)
Nitrogen analysis (optional)
Applications
multi N/C 3300 duo
The highly sensitive TOC/TNb analyzer with its integrated automated solids analysis function is ideal for contract laboratories that handle a wide range of sample types. High sensitivity across a broad working range makes it suitable for both high-matrix and pure water samples, as well as for less homogeneous solid samples. Robust loop injection technology with reverse rinse and parallel purging enables rapid operation in NPOC mode.
Solids analysis
Nitrogen analysis (optional)
Applications:
multi N/C 3300 HS
The TOC analyzer for applications at the ultra-trace level. By dosing directly from the syringe pump, the device can utilize high sample volumes of up to 3 mL, thereby achieving the highest sensitivity and repeatability when analyzing the lowest concentrations of TOC and TN.
Solids analysis (optional)
Nitrogen analysis (optional, can be retrofitted)
Applications:
multi N/C 3300 – Flow injection TOC analyzer
The multi-talented device for high throughput laboratories. Robust loop injection technology with reverse rinse and parallel purging for rapid operation in NPOC mode. Also suitable for pharmaceutical applications such as TOC cleaning validation.
Order number: 450-500.500-2
multi N/C 3300 – Flow injection TOC analyzer
The multi-talented device for high throughput laboratories. Robust loop injection technology with reverse rinse and parallel purging for rapid operation in NPOC mode. Also suitable for pharmaceutical applications such as TOC cleaning validation.
Order number: 450-500.500-2
multi N/C 3300 duo – Fully automated TOC analyzer for liquids and solids
The fast all-rounder for liquid and solid samples. Combination of flow injection with intelligent rinsing and injection technology for particulate water samples with catalyst-free sample oxidation in a ceramic combustion tube for solids.
Order number: 450-500.504-2
multi N/C 3300 duo – Fully automated TOC analyzer for liquids and solids
The fast all-rounder for liquid and solid samples. Combination of flow injection with intelligent rinsing and injection technology for particulate water samples with catalyst-free sample oxidation in a ceramic combustion tube for solids.
Order number: 450-500.504-2
multi N/C 3300 HS – TOC/TN analyzer for pharmaceutical applications
Highly sensitive TOC/TN trace analysis with high-temperature catalytic combustion. Extremely precise in the ultra-trace range thanks to high injection volumes. Module for direct swab combustion is available for TOC cleaning validation.
Order number: 450-500.600-2
multi N/C 3300 HS – TOC/TN analyzer for pharmaceutical applications
Highly sensitive TOC/TN trace analysis with high-temperature catalytic combustion. Extremely precise in the ultra-trace range thanks to high injection volumes. Module for direct swab combustion is available for TOC cleaning validation.
Order number: 450-500.600-2
Applications
Applications
-
Determination of Organic Impurities (TOC) In Raw Materials for Cathode Production in Lithium-Ion Batteries (EN)
-
Bestimmung organischer Verunreinigungen (TOC) in Rohstoffen für die Kathodenherstellung für Lithium-Ionen Batterien (LIB) (DE)
-
Nutzen der TIC-Analytik beim Lithium-Akku-Recycling zur Lithium-Rückgewinnung (DE)
-
Benefits of TIC analysis in lithium battery recycling for Lithium recovery (EN)
-
原油の脱塩工程に伴うブラインサンプルのTOC測定 (JP)
-
DIN EN ISO 20236に準拠した海水中のTOC測定 (JP)
-
TOC Determination in Brine Samples coming from Desalting Process (EN)
-
TOC-Bestimmung in Sole-Proben aus der Erdöl-Entsalzung (DE)